Shimadzu Spd20a User Manual

If your baseline is drifting, it’s often dirty cell lenses rather than the lamp itself. You can dismantle the flow cell to clean or replace these lenses.

| Symptom | Probable Cause | Corrective Action | |---------|----------------|--------------------| | High baseline noise | Air bubbles in cell | Flush with degassed mobile phase or remove cell and rinse. | | Negative peaks | Sample absorbs less than mobile phase at set wavelength | Change wavelength or check mobile phase UV cutoff. | | Lamp fails to ignite | Lamp expired or power supply issue | Check lamp hours; replace D2 lamp. | | No signal at high sensitivity | Flow cell window dirty | Clean cell with 1M HNO₃, then water, then methanol. | | Wavelength error | Stepper motor failure or filter stuck | Run Wavelength Calibration from service menu (qualified personnel only). | shimadzu spd20a user manual

The is a cornerstone of modern high-performance liquid chromatography (HPLC). As a UV/Vis spectrophotometric detector, it is renowned for its sensitivity, dual-wavelength capability, and exceptional noise reduction. However, to truly leverage its power, one must master its documentation: the Shimadzu SPD-20A user manual . If your baseline is drifting, it’s often dirty